For micro and delicate instruments, use manual cleaning(stepC). Instruments should be processed in a cleaner for the full recommended cycle time – usually 5 to 10 minutes.
Place instruments in an open position in the ultrasonic cleaner. Make sure that “Sharp”(scissors, knives osteotomes, etc.) blades do not touch other instruments.
All Instruments have to be fully submerged.Do not place dissimilar metals (stainless, copper, chrome plated, etc.) in the same cleaning cycle.
Change solution frequently – at least as often as the manufacturer recommends.Rinse instruments after ultrasonic cleaning with water to remove the ultrasonic cleaning solution.
For micro and delicate instruments, use manual cleaning(stepC). Instruments should be processed in a cleaner for the full recommended cycle time – usually 5 to 10 minutes.
Place instruments in an open position in the ultrasonic cleaner. Make sure that “Sharp”(scissors, knives osteotomes, etc.) blades do not touch other instruments.
All Instruments have to be fully submerged.Do not place dissimilar metals (stainless, copper, chrome plated, etc.) in the same cleaning cycle.
Change solution frequently – at least as often as the manufacturer recommends.Rinse instruments after ultrasonic cleaning with water to remove the ultrasonic cleaning solution.
Follow our recommendations but ensure instruments are lubricated after the last rinse cycle and before the sterilisation cycle.
Most instrument manufacturers recommend ultrasonic cleaning as the best and most effective way to clean surgical instruments, particularly those with hinges, locks and
other moving parts. If ultra-sonic cleaning is not available observe the following steps.
I. Use stiff plastic cleaning brushes (nylon, etc.) Do not use steel wool or wire brushes except specially recommended stainless steel wire brushes for instruments such as bone
files, or on stained areas in knurled handles.
II. Use only neutral PH(7) detergents because if not rinsed off properly, low PH detergents will cause breakdown of stainless protective surface and black staining. High PH detergent will cause surface deposit of brown stain, which will also interfere with smooth operation of the instrument.
III. Brush delicate instruments carefully and, if possible, handle them totally separate from general instruments.
IV. Make sure all instrument surfaces are visibly clean and free from stains and tissue. This is a good time to inspect each instrument for proper function and condition. Check
and make sure that Scissors blades glide smoothly all the way (they must not be loose when in closed position). Test scissors by cutting into thin gauze. Three quarters of the
length of blade should cut all the way to the scissor tips, and not hang up. Forceps (pickups) have properly aligned tips. Hemostats and Needle Holders do not show light
between the jaws, lock and unlock easily, joints are not too loose. Check Needle Holders for wear on jaw surfaces. Suction tubes are clean inside. Retractors function properly.
Cutting instruments and knives have sharp, undamaged blades.
V. After scrubbing, rinse instruments thoroughly under running water. While rinsing, open and close Scissors, Hemostats, Needle Holders and other hinged instruments to
make sure the hinge areas are rinsed out, as well as the outside of the instruments.
Most instrument manufacturers recommend ultrasonic cleaning as the best and most effective way to clean surgical instruments, particularly those with hinges, locks and
other moving parts. If ultra-sonic cleaning is not available observe the following steps.
I. Use stiff plastic cleaning brushes (nylon, etc.) Do not use steel wool or wire brushes except specially recommended stainless steel wire brushes for instruments such as bone
files, or on stained areas in knurled handles.
II. Use only neutral PH(7) detergents because if not rinsed off properly, low PH detergents will cause breakdown of stainless protective surface and black staining. High PH detergent will cause surface deposit of brown stain, which will also interfere with smooth operation of the instrument.
III. Brush delicate instruments carefully and, if possible, handle them totally separate from general instruments.
IV. Make sure all instrument surfaces are visibly clean and free from stains and tissue. This is a good time to inspect each instrument for proper function and condition. Check
and make sure that Scissors blades glide smoothly all the way (they must not be loose when in closed position). Test scissors by cutting into thin gauze. Three quarters of the
length of blade should cut all the way to the scissor tips, and not hang up. Forceps (pickups) have properly aligned tips. Hemostats and Needle Holders do not show light
between the jaws, lock and unlock easily, joints are not too loose. Check Needle Holders for wear on jaw surfaces. Suction tubes are clean inside. Retractors function properly.
Cutting instruments and knives have sharp, undamaged blades.
V. After scrubbing, rinse instruments thoroughly under running water. While rinsing, open and close Scissors, Hemostats, Needle Holders and other hinged instruments to
make sure the hinge areas are rinsed out, as well as the outside of the instruments.
If instruments are to be reused or autoclaved:
A. Lubricate all instruments which have any “metal to metal” action such as scissors, hemostats, needle holders, self-retaining retractors, etc. Recommend surgical lubricants such as instrument milk are best. Do not use WD-40, oil or other industrial lubricants.
B. Put instruments up for autoclaving either individually or in sets.
Most cold sterilization solutions render instruments sterile only after a 10 hour immersion. This prolonged chemical action can be more detrimental to the surgical instruments than the usual 20 minute autoclave cycle. If the instruments need to be “disinfected” only, cold sterilization is okay as disinfection will place in only 10 minutes. But keep in mind the difference between:
Always use the proper sterilization/cleaning technique to render the instrument in required condition for use. For instruments with Tungsten Carbide inserts (Needle Holders, Scissors, Tissue Forceps), we do not recommend use of solutions containing Benzyl Ammonium Chloride which will destroy the Tungsten Carbide Inserts.
Store sterilized items in a clean, dry place that is protected from dust, dirt, and moisture. Sterile items must be stored off the floor. Handling increases the chances of punctures of sterilized bags. Sterilized items must be stored separately from dirty equipment/instruments.
Unlock all instruments and sterilize them in an open position. Place heavy instruments on bottom of set (when two layers are required). Never lock an instrument during autoclaving. It will not be sterile as steam cannot reach the metal to metal surfaces. The instrument will develop cracks in hinge areas because of heat expansion during the autoclave cycle.
Disposable paper or plastic pouches are ideal. Make sure you use a wide enough pouch (4″ or wider) for instruments with ratchet locks such as needle holders and hemostats so the instrument can be sterilized in an open (unlocked) position.
At the end of the autoclave cycle – before the drying cycle – unlock autoclave door and open it more than a crack (about 3/4″).
Then run dry cycle for the period recommended by the autoclave manufacturer. If the autoclave door is opened fully before the drying cycle, cold room air will rush into the chamber, causing condensation on the instruments. This will result in water stains on instruments and also cause wet packs. If you have any unusual staining on your instruments during sterilization, contact your local instrument representative.